6. Push the „SET”button to select the second phase of the program. You can program up to
5 phases of your own in each of the three congurable programs. If you want to reduce a
congured program of 5 phases to less phases, you simply set the time of the rst unwanted
phase to zero.
7. The programming is completed by pressing the „PRG” button and the program is stored.
Selectable frequency and pulse width settings:
Continuous-Mode: 1 to 200 Hz, 50 to 450 µS
Burst-Mode: 35 to 200 Hz, 50 to 200 µS
Modulated-Mode: 1 to 200 Hz, 50 to 450 µS
Work/Rest-Mode: 1 to 200 Hz, 50 to 450 µS
It is not possible to program more than 300 µS if the frequency is above 100 Hz.
The settings can be made as follows.
Frequency: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 12, 14, 15 Hz, and then increasing in 5 Hz steps until
max. 200 Hz
Pulse width; 50 to 450 µS in steps of 10 microseconds
Intensity: 50-100% in steps of 10%
Modulation time: 2, 3, 4, 5, 6, 7, 8, 9, 10, 12, 15, 20, 30, 40, 50, 60 seconds
Work time: 2-99 seconds in steps of 1 second
Ramp up time: 0,1-9,9 seconds in steps of 1 second
Ramp down time: 0,1-9,9 seconds in steps of 1 second
Rest time: 2-99 seconds in steps of 1 second
Delay: 0,0-5,0 seconds in steps of 0,1 second
Session time: 1-599 minutes in steps of 1 minute