38
Setting the
Custom Programs 13-15
1. Switch on DoloBravo with „ON/OFF” button.
2. Choose the desired program (13-15) with „PRG” button.
3. Press „SET” button for at least ve seconds. DoloBravo changes to the programming mode
and the upper watch, which indicates phase duration, ashes to show that this function can be
altered.
In the following please use the left „+” button always then, if you want to store your
choice and to change to the next step. The right „+/-” buttons are respectively used in
order to alter the chosen and ashing parameter.
4. Select with the right „+/-”buttons the desired time for the rst phase.
5. By pressing the left „+”button you will be able to store your choice and change to the next
step. In the display the mode now ashes, which you adjust according to your wishes by
pushing the right „+/-”buttons.
If you have chosen „CONT” or „BURST” you can determine frequency and after storage and
change with the left +button respectively the pulse width by pushing the right „+/-” buttons.
If you have chosen „WORK/REST”- rhythm you can determine the following parameters one
after the other by pushing the right „+/-”buttons.
- frequency (Hz)
- pulse width (µS)
- work time (sec)
- rest time (sec = recovery time)
- increase / decrease ramp up time of the pulse (RU)
- increase / decrease ramp down time of the pulse (Rd)
- synchronous (SY) or alternating (AL) output of the channels A and B. If you have selected
“SY” you can program delay-time (dL) of 0 sec. up to 5 sec. from channel A to B.
If you have chosen „MOD” you can alter the following parameters one after the other.
- low limit of frequency (Hz LO)
- upper limit of frequency (Hz HI)
- low limit of pulse width ( µS LO)
- upper limit of pulse width (µS HI)
- low limit of intensity (AdJ) - the display in the right corner below shows values from 0.5 to
1.0. This adjusts the intensity of current within the modulation time between 50% (=0.5) and
100% (=1.0) of the original set mA-value. The intensity, of a modulated program, can never
increase above the 100% of the set value.
- modulation time (possible choice: 2 sec. up to 60 sec)