Guidance and manufacturer‘s declaration - electromagnetic emissions
The DoloBravo unit is intended for use in the electromagnetic environment specied below.
The customer or the user of the DoloBravo should ensure that it is used in such an
environment.
Emissions test Compliance
Electromagnetic environment
guidance
RF emissions
CISPR 11
Group 1
The DoloBravo uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby electro-
nic equipment.
RF emissions
CISPR 11
Class B
The DoloBravo is suitable for use in all
establishments, including domestic esta-
blishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not Applicable
Voltage uctuations/
icker emissions
IEC 61000-3-3
Not Applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
The DoloBravo unit is intended for use in the electromagnetic environment specied below.
The customer or the user of the DoloBravo should ensure that it is used in such an environ-
ment.
Immunity test
IEC 60601-1-2
test level
Compliance
level
Electromagnetic environ-
ment guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/-6 kV contact
+/-8 kV air
+/-6 kV contact
+/-8 kV air
Floors should be wood, con-
crete or ceramic tile. If oors
are covered with synthetic
material, the relative humidi-
ty should be at least 30 %.
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic
elds should be at levels
characteristic of a typical
location in a typical commer-
cial or hospital environment.