SineBravo Operators Manual
31
Electromagnetic Compatibility
Guidance and manufacturer’s declaration - electromagnetic Immunity
The SineBravo is intended for use in the electromagnetic environment specied below. The customer or the user
of the SineBravo should assure that it is used in such an environment
Radiated RF
IEC61000-4-3
(Test speci-
cations for
ENCLOSURE
PORT
IMMUNITY
to RF wireless
communications
equipment)
Test
Frequency
(MHz)
Band
(MHz)
Service Modulation
Modu-
lation
(W)
Dis-
tan-
ce
(m)
IMMUNITY
TEST
LEVEL
(V/m)
385
380
–390
TETRA 400
Pulse
modulation
18 Hz
1,8 0,3 27
450
380
–390
GMRS 460,
FRS 460
FM c)
± 5 kHz
deviation
1 kHz sine
2 0,3 28
710
704 –
787
LTE Band 13,
17
Pulse
modulation
217 Hz
0,2 0,3 9
745
780
810
800 –
960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
18 Hz
2 0,3 28
870
930
1720
1 700
1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band
1, 3,
4, 25; UMTS
Pulse
modulation
217 Hz
2 0,3 28
1845
1970
2450
2 400
2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
217 Hz
2 0,3 28
5240
5 100
5 800
WLAN 802.11
a/n
Pulse
modulation
217 Hz
0,2 0,3 9
5240
5785

Related product manuals