ELT-400 7 Remote control
7-9
SET:DETECTOR<SP><param>[CR]<LF>
SET:DETECTOR?[CR]<LF>
Response:
<param><CR><LF>
Meaning:
Selects the detection mode.
param Explanation
RMS The RMS value measured using the selected operating mode
will be output.
PEAK The peak value measured using the selected operating mode
will be output.
STND In “Exposure STD” mode only:
The detection mode is selected automatically according to the
selected standard.
Table 7-6 Detection mode selection settings