7 Remote control ELT-400
7-14
MEAS
Meaning:
Transfers the measured values and, where appropriate, the status
information.
Different modes for transferring individual values and for transferring values
continuously are provided.
Remote control
command
Explanation
MEAS?[CR]<LF> Transfers the latest measured value / status on
request.
Asynchronous query:
The same measured value is transferred several
times if the query is made more often than the
measurement value is updated
(measured value is refreshed 4 times per second).
MEAS:ARRAY?<SP>
<param>[CR]<LF>
Consecutive transfer of a selectable number of
measured values / statuses at regular intervals
(measured value is refreshed 4 times per second).
Synchronous query:
Each measured value is transferred once only.
The transfer of the first measured value may be
delayed slightly.
param: Number of measured values requested
(setting range: 1 to 65535)
Transfer can be terminated prematurely using the
command MEAS:STOP.
MEAS:START[CR]<LF> Continuous transfer of measured values / statuses at
regular intervals (measured value is refreshed 4 times
per second).
Synchronous query:
Each measured value is transferred once only.
The transfer of the first measured value may be
delayed slightly.
Transfer is stopped with the command
MEAS:STOP.
MEAS:STOP[CR]<LF> Stops continuous transfer