EasyManuals Logo
Home>National Instruments>Controller>cDAQ-9133

National Instruments cDAQ-9133 User Manual

National Instruments cDAQ-9133
145 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #118 background imageLoading...
Page #118 background image
© National Instruments | 5-35
NI cDAQ-9132/9133/9134/9135/9136/9137 User Manual
Routing a Signal to Counter n Aux
Each counter has independent input selectors for the Counter n Aux signal. Any of the following
signals can be routed to the Counter n Aux input:
Any PFI terminal
AI Reference Trigger
AI Start Trigger
Analog Comparison Event
Change Detection Event
In addition, a counters Internal Output, Gate or Source can be routed to a different counters
Aux. A counter’s own gate can also be routed to its Aux input.
Some of these options may not be available in some driver software. Refer to the Device Routing
in MAX topic in the NI-DAQmx Help or the LabVIEW Help for more information about available
routing options.
Counter n A, Counter n B, and Counter n Z Signals
Counter n B can control the direction of counting in edge counting applications. Use the A, B,
and Z inputs to each counter when measuring quadrature encoders or measuring two pulse
encoders.
Routing Signals to A, B, and Z Counter Inputs
Each counter has independent input selectors for each of the A, B, and Z inputs. Any of the
following signals can be routed to each input:
Any PFI terminal
Analog Comparison Event
Routing Counter n Z Signal to an Output Terminal
You can route Counter n Z out to any PFI terminal.
Counter n Up_Down Signal
Counter n Up_Down is another name for the Counter n B signal.
Counter n HW Arm Signal
The Counter n HW Arm signal enables a counter to begin an input or output function.
To begin any counter input or output function, you must first enable, or arm, the counter. In some
applications, such as a buffered edge count, the counter begins counting when it is armed. In
other applications, such as single pulse-width measurement, the counter begins waiting for the
Gate signal when it is armed. Counter output operations can use the arm signal in addition to a
start trigger.

Table of Contents

Other manuals for National Instruments cDAQ-9133