Calibration Procedure

Other manuals for National Instruments PXIe-4467

Questions and Answers

  • V
    villegaschristopherAug 11, 2025
    How to fix National Instruments Control Unit PXIe-446x not appearing in MAX?
    • E
      Erin Baker MDAug 11, 2025
      If the National Instruments Control Unit PXIe-446x module is not showing up in Measurement & Automation Explorer (MAX), try the following: * In MAX, expand 'Devices and Interfaces'. * Expand the 'Chassis' tree and refresh the list by pressing F5. * If it's still not listed, power off the system, ensure the hardware is correctly installed, and restart. * Check the Device Manager to verify the PXIe-446x appears there under an NI entry. If there are error conditions, reinstall the NI-DAQmx driver.
  • J
    Jessica SmithAug 21, 2025
    What to do if the ACCESS LED is off when the National Instruments Control Unit chassis is on?
    • E
      Elizabeth WhitakerAug 21, 2025
      If the ACCESS LED is off when the National Instruments Control Unit chassis is on, it could indicate a problem with the chassis power rails, the hardware module, or the LED itself. First, verify that the PXIe-446x appears in MAX. Then: 1. Disconnect any signals from the module front panels. 2. Power off the chassis. 3. Remove the module and inspect it for damage. Do not reinstall a damaged module. 4. Install the module in a different chassis slot. 5. Power on the chassis again. 6. Check if the module appears in MAX. Reset the module in MAX and perform a self-test.
  • B
    Brent KirkSep 1, 2025
    What to do if National Instruments Control Unit PXIe-446x fails the self-test?
    • A
      angel48Sep 1, 2025
      If the National Instruments Control Unit PXIe-446x fails the self-test, try the following steps: 1. Restart the system. 2. Launch MAX and perform the self-test again. 3. Power off the chassis. 4. Reinstall the failed module in a different slot. 5. Power on the chassis. 6. Perform the self-test again.

National Instruments PXIe-4467 Specifications

General IconGeneral
Form FactorPXIe
Number of Analog Output Channels2
Input TypeDifferential
Sampling RateUp to 204.8 kS/s
Resolution24 bits
Output Range±10 V
Operating Temperature0 °C to 55 °C
Input CouplingAC/DC
Output CouplingDC
Input Range±10 V

Related product manuals