PXIe-4466/4467 Calibration Procedure
40 ni.com/services
1. Open a calibration session using the DAQmx Initialize External Calibration
VI with the following parameters:
deviceName: Dev_name
password: NI
calHandle: &myCalHandle
2. Close the session with the DAQmx Close External Calibration VI with the
following parameters:
calHandle: myCalHandle
action: DAQmx_Val_Action_Commit
Revision History
Revision Section Changes
378608A-01
February 2022
This is the initial release
version of the PXIe-4466/4467
Calibration Procedure.