EasyManua.ls Logo

National Instruments PXIe-4467 - Timebase Frequency Accuracy Verification

National Instruments PXIe-4467
41 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PXIe-4466/4467 Calibration Procedure
34 ni.com/services
Timebase Frequency Accuracy Verification
Test Limits
Table 26: Timebase Frequency Accuracy Limits
Function Generator
Output
As-Found Test Limit As-Left Test Limit
Anplitude
(V
rms
)
Frequency
(kHz)
Lower Limit
(kHz)
Upper Limit
(kHz)
Lower
Limit (kHz)
Upper Limit
(kHz)
6.3 10.000 9.99973 10.00027 9.99992 10.00008

Other manuals for National Instruments PXIe-4467

Related product manuals