EasyManua.ls Logo

National Instruments PXIe-4467 - Perform Verification; AI Offset Accuracy Verification

National Instruments PXIe-4467
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PXIe-4466/4467 Calibration Procedure
© National Instruments Corporation 9
1. Launch MAX.
2. Select My System»Devices and Interfaces»NI PXIe-4466 or My
System»Devices and Interfaces»NI PXIe-4467.
3. Start self-calibration using one of the following methods:
4. Click Self-Calibrate in the upper right corner of MAX.
5. Right-click the name of the DUT in the MAX configuration tree and select
Self-Calibrate from the drop-down menu.
Perform Verification
AI Offset Accuracy Verification
Test Limits
Table 1: AI Offset Verification Limits
Device
Gain (dB)
As-Found Test Limit As-Left Test Limit
Lower Limit
(mV)
Upper Limit
(mV)
Lower Limit
(mV)
Upper Limit
(mV)
DC Coupled
-20 -5.0 5.0 -0.9 0.9
-10 -2.0 2.0 -0.3 0.3
0 -0.5 0.5 -0.1 0.1
10 -0.2 0.2 -0.05 0.05
20 -0.1 0.1 -0.04 0.04
30 -0.1 0.1 -0.04 0.04

Other manuals for National Instruments PXIe-4467

Related product manuals