NOTE: S1-E must be in, except on I02 range.
NOTE: When using the default (S4-B selecting the 4 MSD), divide display
counts by 100 before using Table 4-3.
4.4 Offset Range Selection
Using the same LO, UO, LD, and UD numbers, calculate the zero offset
using the appropriate formula below:
LD x (UO - LO)
V01 & I01 Offset =
(UD - LD)
LD x (UO - LO)
LD x (UO - LO)
I02 Offset =
(UD - LD)
+ 1
Select one of the ranges from the Offset Table (see Table 4-4) which
includes the calculated offset. Then reposition push-on jumpers S1-A, S1-
B, S1-C, and S1-D as required.
VOLTAGE
RANGE
COUNTS/VOLT
CURRENT
RANGE
COUNTS/MA
S2
VR1
48.0 - 63.0 CR1 27.2 - 35.7
CF
VR2
62.5 - 82.0 CR2 35.4 - 46.5
DF
VR3
81.5 - 106.5 CR3 46.2 - 60.3
E
VR4 106.0 - 138.5 CR4 60.1 - 78.5
BE
VR5 138.0 - 180.0 CR5 78.2 - 102.0
AF
VR6 179.5 - 234.0 CR6 101.7 - 132.6
CE
VR7 233.5 - 304.0 CR7 132.3 - 172.2
BDE
VR8 303.5 - 395.0 CR8 171.9 - 223.8
--
VR9 394.5 - 513.5 CR9 223.5 - 290.9
AE
VR10 513.5 - 667.5 CR10 290.9 - 378.2
C
VR11 667.0 - 867.5 CR11 377.9 - 491.5
BD
*VR12 867.0 - 1127.5 CR12 491.2 - 638.8
A
* Default Setting
Table 4-3 Span Ranges