December 2019  105  LDC-3706 Series Laser Controller 
 
9.  The instrument will ramp to a new setpoint. Repeat step 7 for the newly calculated 
photodiode current at this setpoint. 
10. Wait for the 7-segment display to exit the calibration mode.  Once the self-calibration is 
completed, the calibration constants will be stored to the non-volatile memory and the 
display will return to its previous state.