Remote Command Set 75
Description Indefinite move command.
Syntax xxMVnn
Remarks This command is used to move an axis indefinitely. If this command is issued when an
axis’ motion is in progress, the controller will ignore this command and generate
“MOTION IN PROGRESS” error message.
Issue a Stop (ST) or Abort (AB) motion command to terminate motion initiated by MV
Argument Value Description
xx 1 to 4 Axis number
nn + Move desired axis in positive direction
- Move desired axis in negative direction
Example
1MV+ (Move motor 1 indefinitely in positive direction)
AB (Abort /halt motion)
2MV- (Move motor 2 indefinitely in negative direction)
2ST (Stop motion)
See Also
PA, PR, ST, AB