Pyrocam IV Operator’s Manual Document No. 50337-001 Rev F 2/21/2017 Page 13
Read Pixel List
Click to read the current pixel
List
from the camera.
Clear Pixel List
Clears all marked pixels from the correction
List.
Software Bad Pixel Correction
Enable/Disable the bad pixel correction simulation built into the BeamGage software. This
allows the operator to preview the effect of the current BP
List
before writing it into the
Pyrocam’s EEPROM.
Bad Pixel Correction Map
Toggle control to show/hide the Marked pixels currently entered in the correction
List
.
Pixel Selection Type
Select either a single pixel, row of pixels, or column of pixels when adding/removing pixels
from the correction
List
.
Pixel List
You can view and select a pixel to edit in the pixel
List
using this dropdown control. It
contains all pixel coordinates currently entered in the
List.
Gain Correction
Click this control to enable/disable gain correction in the Pyrocam. Gain Correction (GC)
will default to the ON state whenever a GC table is loaded into the Pyrocam. If this control
is deactivated, it indicates that the Pyrocam does not have a gain correction table
programmed in its EEPROM. The procedure for creating a Gain Table is described in
APPENDIX D.
Note: The Pyrocam has separate gain correction tables programmable for pulsed and
chopped operation. Many of the Pyrocams will ship with a valid 50Hz chopped gain
correction table installed at the factory. Pyrocams will almost never ship with a
pulsed gain correction table. Factory generated GC files are supplied with the
Pyrocam in case the programmed table is accidentally deleted or corrupted.