1.6. Check scope of delivery
• Open the packaging with care. Avoid fingerprints and sweat on the camera lens. Pre-
vent the camera and some accessories from falling and being damaged.
• Remove all components from the packaging and check for completeness according to
the delivery note. The microscope could be packaged in more than one carton depend-
ing on the model or accessories.
• Handle the microscope with care and always protect it against violent impact and
vibrations.
• Keep the original packaging for possible longer storage or return of the device to the
supplier in case a repair or revision is needed.
Never lift the microscope by the cross stage, trinocular viewing head or coarse/
fine focusing knobs. This can damage the microscope!
• Always carry the microscope with both hands.
• On the upper part of the microscope body there is a carrying handle.
• The microscope back plate is designed with a holing device to effec-
tively store the long power cord, improve the cleanliness of the labo-
ratory, and reduce the tripping accident caused by the long power
cord during the carrying process.
1.7. Site location
• Select a suitable location before setting up the microscope.
• Place the microscope on a resistant surface away from strong sunlight, heat sources,
high humidity, high dust exposure as well as strong vibration.
Do not expose the device to temperatures below 5°C or higher than +30°C! If
the device is exposed to temperatures outside this range, this will cause irreparable
damage to optical and/or mechanical parts which are not covered by warranty.
› Working temperature: 0°C℃~40°C℃;
› Maximum relative humidity: it is 80% for temperature of 31°C℃, the following is the
linear decrease, 70% for 34°C℃, 60% for 37°C℃, 50% for 40°C℃;
› The highest elevation is 2000m
› Storage and transport environment:
› Range of working temperature: -25°C℃~+65°C℃
› Range of relative humidity: 0%~90%
• Place the microscope on a surface that meets the following criteria:
› Flat / Level
› Vibration-free
› Hard surface and non-flammable
› Chemical and mechanical resistant
• Place the microscope so that:
› The ventilation slits on the underside are not blocked
› There is at least 10 cm clearance around the device or to other devices or the wall.
› You can disconnect it from the mains at any time
• The mains cable must be freely accessible at all times, as the mains cable is intended
as a disconnecting device from the mains.
• Make sure the residual moisture is fully eliminated before use.
• For use in warm and humid climates, all optical components of the microscope are
already equipped with protection against fungus infestation.
IMPORTANT NOTES