Important Information (Precautions)
Battery change and battery cover
Do not operate the stimulator with the battery cover removed. Disconnect all the leads prior
to opening the battery cover. The clinician/carer should not be in contact with the user when
changing the battery.
Equipment modication
Do not tamper with or modify the equipment.
Shelf life
Do not use the consumables if they have passed their expiry date.
Equipment re-use
Electrodes, elasticated tubular stockings and insoles are single patient use only. If re-using
other parts of the system ensure a thorough clean using anti-bacterial wipes.
Storage temperature
Do not store outside of the temperature limits as detailed in the specication on page 47.
If the ODFS
®
Pace (XL) or accessories (excluding electrodes) have been stored at the
minimum (-20°C) or maximum (70°C) storage temperatures, allow at least 10 minutes in
ambient temperature before use. This time may be required following system removal from
cold storage for the screen on the stimulator to reach full clarity.
Use of small electrodes
The use of small electrodes (less than 32mm diameter) may result in higher charge density
especially when using high output settings (current, pulse width, frequency) and may lead to
an increase in the likelihood of skin irritation. When possible, use larger electrodes. Always
monitor skin condition closely.
Hot surfaces
The surface of the ODFS
®
Pace (XL) can reach a maximum temperature of 44°C when using
the device at very high output levels and in a maximum ambient temperature of 40°C. Do
not wear next to the skin if using very high output levels with an ambient temperature above
36°C, as this may cause burning of the skin.