EasyManuals Logo

NIDEC TOSOK DAG2000 User Manual

NIDEC TOSOK DAG2000
33 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
DIGITAL AIR MICROMETER USERS MANUAL (Second Edition)
9
3.3 For first time measurement tool connection cases
Shown here is the procedure up to measurement wherein measurement tool is connected to the
main body for the first time.
Note: There are cases wherein detector adjustment is not needed depending on the measurement
tool.
Turn ON compressed air
Turn ON power
Switch to settings mode
Change the settings
Switch to master
calibration mode
(detector adjustment)
Detector adjustment
Switch to master
calibration mode (master
calibration)
Master calibration
Measurement
Switch from measurement mode to settings mode
Because of work tolerance change, change the judgmen
t
limit values.
Settings value change can be done in setting mode.
Please refer [Chapter 6, 6.4 Settings] for details
At settings mode, input the settings value for the
measurement tool that will be connected.
Please refer to [Chapter 6, 6.4 Settings] for details
Switch to measurement mode after finishing settings
change
Please refer to [Chapter 6, 6.4 Settings] for details
Switch from measurement mode to detector adjustment
Please refer to [Chapter 6, 6.5 Detector Adjustment] fo
details
Switches from detector adjustment to master calibration
Refer to [Chapter 6, 6.5 Detector Adjustment] fo
details.
Adjustment of detector through ZERO and sensitivity
tuner.
Please refer to [Chapter 6, 6.5 Detector Adjustment] fo
r
details

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NIDEC TOSOK DAG2000 and is the answer not in the manual?

NIDEC TOSOK DAG2000 Specifications

General IconGeneral
BrandNIDEC TOSOK
ModelDAG2000
CategoryMeasuring Instruments
LanguageEnglish