Chapter 1 Microscopy Procedures
4
Chap. 1-1
Microscopy Procedures Bright-field Microscopy
2
Select the setup target with the toggle pattern switch, and then switch to the setting mode to
configure the motorized functions.
Before using a motorized swing-out condenser for the first
time, configure the swing-out of the top lens. (→See
Chapter 2, Section 13.1 for details.)
To use the toggle function, configure the toggle pattern.
(→See Chapter 2, Section 12.2 for details.)
3
Set the setting mode switch to the left side (operation mode).
Set the setting mode switch on the remote control pad to
the left side.
OFF
TOGGLE
PATTERN
MEMORY
12
4
Turn on the dia-illumination LED.
Press the dia-illumination LED ON/OFF switch to turn on
the LED.
P
O
WER
Dia-illumination LED switch ON
5
Lower the condenser slightly from the uppermost position.
Turn the condenser focus knob until the condenser is
positioned at the upper limit (where it clicks to a stop), and
then lower it a little.
CLAMP
TORQUE
0.1
0.3
0.4
0.5
0.2
0.9
0.8
Lower the condenser from upper limit.