Chapter 8 Troubleshooting Tables
3 Electrical Problems
83
8
3
Electrical Problems
Lamp does not light when switched on.
Causes Corrective Measures
No electrical power.
Check power cord connection. (P.69)
Lamp bulb not inserted.
Insert correctly. (P.72)
Lamp bulb burnt out.
Replace bulb. (P.72)
Incorrect lamp used.
Use the specified lamp. (P.72)
Fuse blown out.
Replace fuse. (P.73)
Flickering or unstable lamp brightness.
Causes Corrective Measures
Lamp bulb about to fail.
Replace bulb. (P.72)
Power cord not correctly connected.
Connect correctly.
Bulb not correctly inserted into socket.
Insert correctly. (P.72)
Sudden lamp failure.
Causes Corrective Measures
Incorrect lamp used.
Use the specified lamp. (P.72)
Incorrect input voltage.
Select the voltage that corresponds to
the voltage provided in your region
using the voltage selector. (P.68)

Other manuals for Nikon eclipse E200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nikon eclipse E200 and is the answer not in the manual?

Nikon eclipse E200 Specifications

General IconGeneral
TypeBiological Microscope
Optical SystemCFI60 Infinity Optical System
Objectives4x, 10x, 40x, 100x (Oil)
Observation MethodBrightfield
Magnification40x to 1000x
FocusCoaxial coarse/fine focus knobs
Eyepieces10x
CondenserAbbe condenser
Power SupplyAC 100-240V
Eyepiece TubeBinocular Tube, Inclination 30°

Related product manuals