Installing
the
Condenser
O
Installation
1-1
Turn
the
condenser
focus
knob
to
lower
the
condenser
holder
as
far
as
it
will
go.
1-2
Insert
the
condenser
in
the
condenser
holder.
1-3
Tighten
the
clamp
screw
on
the
left
side
with
the
aperture
scale
on
the
condenser
facing
to
the
front.
1-4
Turn
the
condenser
focus
knob
to
raise
the
condenser
as
far
as
it
will
go.
©
Focusing
and
centering
Focus
and
center
the
condenser
so
that
the
light
passing
through
the
condenser
forms
the
image
at
the
proper
position
on
the
specimen
(i.e.,
at
the
center
of
the
optical
path).
2-1
Focus
on
the
specimen
with
the
10x
objective.
2-2
Close
the
field
diaphragm
to
its
minimum
setting.
2-3
Turn
the
condenser
focus
knob
so
that
the
image
of
the
field
diaphragm
forms
on
the
specimen.
2-4
Adjust
condenser
centering
screws
so
that
the
image
of
the
field
diaphragm
appears
roughly
at
the
center
of
the
eyepiece
view
field.
2-5
Move
the
40x
objective
into
the
optical
path.
Turn
the
fine
focus
knob
to
focus
on
the
specimen.
2-6
Turn
the
condenser
focus
knob
so
that
the
image
of
the
field
diaphragm
forms
on
the
specimen.
2-7
Adjust
the
condenser
centering
screws
so
that
the
image
of
the
field
diaphragm
appears
at
the
center
of
the
eyepiece
view
field.
This
adjustment
is
easier
to
make
if
you
adjust
the
size
of
the
field
diaphragm
to
slightly
smaller
than
the
eyepiece
view
field.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nikon Eclipse E400 and is the answer not in the manual?

Nikon Eclipse E400 Specifications

General IconGeneral
TypeBiological Microscope
Optical SystemCFI60 Infinity Optical System
NosepieceQuadruple nosepiece
Magnification40x to 1000x
Power SupplyAC 100-240V, 50/60Hz
Lamp Power6V 30W
Eyepiece10x
Objective Lenses4x, 10x, 40x, 100x
StageMechanical stage
FocusCoaxial coarse and fine focusing
CondenserAbbe condenser, N.A. 1.25
Observation TubeBinocular
Stage Dimensions134mm
Interpupillary Distance Adjustment50-75mm
Diopter Adjustment+/- 5 diopters
IlluminationBuilt-in Koehler illumination