Nikon N Series/K Series Total Station Instruction Manual 61
Applications 4
If more than the minimum required data is
available, a standard deviation screen
appears.
To take measurements to strengthen
geometry of the resection, press the
Add
softkey. For information about the
View
softkey, see Advanced feature: Viewing and
deleting a measurement in resection,
page 62.
8. When the results are satisfactory, record the
station. To do this, press
[ENT] or the REC softkey.
9. Enter the height of instrument, if required. Press
[ENT]. The ST field defaults to the last recorded
PT + 1.
10. To change the station name, move to the ST field
and edit or replace the text.
BS defaults to the first observed point.
11. To change the BS, press the
Change softkey.
12. Select the BS point that you want to use and
press
[ENT].
13. To finish the resection setup, move the cursor to
the BS field and press
[ENT].
If you have set Split ST to Yes, the ST field defaults to the last recorded ST value + 1. For more information, see
Others settings, page 121.
The minimum data required for a resection is either three angle shots, or one angle shot and one distance shot.
If you use a distance shot, the distance between the target points must be greater than the measured distance.
Stn-Z is calculated from distance-measured data. If no distances are measured, then Stn-Z is calculated using
angle-only measurements to points with 3D coordinates.

Table of Contents

Other manuals for Nikon N Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nikon N Series and is the answer not in the manual?

Nikon N Series Specifications

General IconGeneral
BrandNikon
ModelN Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals