17
Control Test
We recommend performing Control Tests for
practice before using your Meter for the first time
to test your blood.
Control Tests should be performed:
For practice to ensure your testing technique is good,
Occasionally as you use the vial of Strips,
When opening a new vial of Strips,
If results seem unusually high or low,
If a vial has been le opened or exposed to extreme
heat or cold, or humidity,
If Meter damage is suspected (Meter was dropped,
crushed, wet, etc.)
Note:
ree levels of TRUEtest Glucose Control Solution
are available for Quality Control Testing. It is
important to perform Control Tests with more
than one level of Control to assure your System
is working properly and your testing technique
is good. Contact place of purchase or call for
assistance to obtain Control.
Caution! Ranges printed on Strip vial label are
for Control Test results only and are
not suggested levels for your blood
glucose. Do not drink Glucose Control.
Tre_OB_GutsENG_E4NPD04r22.indd 17 4/24/12 9:53 AM