33
Caring for TRUEresult
Store System (Meter, Glucose Control, Test Strips) in
Carrying Case to protect from liquids, dust and dirt.
Store in a dry place at room temperature (36-86°F).
DO NOT REFRIGERATE OR FREEZE.
Meter Care, Cleaning/Disinfection
Cleaning removes blood and soil, disinfecting
removes infectious agents.
To Clean the Meter (removing blood or soil):
- W
e suggest cleaning the Meter when visibly dirty.
-
Never put Meter in liquids or allow any liquids to enter
the Test Port.
- W
ipe Meter with a clean, lint-free cloth dampened with
70% Isopropyl alcohol.
-
Let Meter air dry thoroughly before using to test.
-
Do not use bleach to clean the Meter.
For assistance contact Customer Care using the phone
number on the cover of the Owner's Booklet.
For Healthcare Professionals, please refer to the Quality
Assurance/Quality Control Manual for instructions on
Meter care, cleaning and disinfection.
Tre_OB_GutsENG_E4NPD04r22.indd 33 4/24/12 9:54 AM