39
System Specifications
Result Range: 20-600 mg/dL
Sample Size: Minimum 0.5 microliter (0.5 µL)
Sample:
Fresh capillary whole blood, venous blood drawn
in EDTA or heparin tubes, or Glucose Control
Test Time: Results in as little as 4 seconds
Result Value: Plasma values
Assay Method: Electrochemical
Power Supply: One 3V lithium battery
#CR2032 (non-rechargeable)
Total power when active
at full battery = 8.6mW
Battery Life: Approximately 2146 tests or 1.5 years
Automatic shut-o: Aer two minutes of non-use
Weight: 1.66 ounces
Size: 3.44” x 2.16” x 0.69”
Memory Size: 500 glucose results
Operating Range (Meter & Test Strips):
Relative Humidity: 10-90% (Non-condensing)
Temperature: 50-104ºF
Hematocrit: 20-60%
Note: Use within specified environmental
conditions only.
Chemical Composition
TRUEtest Test Strips: Glucose dehydrogenase-PQQ
(Acinetobacter calcoaceticus), mediators, buers and
stabilizers.
TRUEtest Glucose Control: Contents: Water,
d-glucose, buers, viscosity enhancing agent, salts,
dye and preservatives.
Tre_OB_GutsENG_E4NPD04r22.indd 39 4/24/12 9:54 AM