37
Messages
Display
Reason
Temperature
Error
Too Cold/
Too Hot
Sample Not
Detected or
Using Wrong
Test Strip
Used Strip,
Strip outside of
vial too long,
Sample on top
of Strip.
Meter
Error
Test Strip
Error
Action
Move Meter and Strips
to area between
50-104ºF; wait 10 minutes
for System to reach
room temperature before
testing.
Retest with new
TRUEtest Test Strip and
larger sample.
Repeat with new Strip.
Make sure sample is
touched to edge of
Strip (not top). If error
persists, call for assistance.
Call for
assistance.
Retest with new Strip.
If error persists,
call for assistance.