7
Limitations
Do not use the Lactate Plus Meter System in the
diagnosis, treatment, or monitoring of disease or other
conditions.
Use only capillary or venous whole blood.
Do not use serum or plasma.
There is no effect of altitudes up to 12,000 feet
(3,658 meters) above sea level.
Testing outside of Lactate Plus System environmental
specications may cause inaccurate results.
Testing of blood sample with Hematocrit outside of
20% - 65% range may cause inaccurate results.
PN62659B1 Lactate Plus IFU EN.indd 7PN62659B1 Lactate Plus IFU EN.indd 7 8/5/22 11:54 AM8/5/22 11:54 AM