66
Appendix
Instructional Notes
1. Inserting a test strip overrides all other modes except
Upload mode.
2. There is no “All segments Screen” in Upload mode.
3. When in Upload mode, no other meter functions are
available, and there is no response by meter to button
presses.
4. The meter times out in 5 minutes in Upload mode.
5. If in setup mode when the test strip is inserted, the me-
ter saves all values entered up to that point and immedi-
ately switches to test mode. Upon exiting test mode the
meter screen goes blank and does not return to setup
mode.
PN62659B1 Lactate Plus IFU EN.indd 66PN62659B1 Lactate Plus IFU EN.indd 66 8/5/22 11:55 AM8/5/22 11:55 AM