Accuracy
Accuracy of the Lactate Plus Meter system (Lactate Plus Mono-
chrome) was assayed at clinical sites by comparing lactate re-
sults obtained by Lactate Plus Meter vs YSI Reference method
from subjects at clinical sites. Blood lactate measurements
obtained from Nova Lactate Plus Meter were compared with
the same subjects’ results from the YSI, a laboratory reference
method. The latter results were plasma-derived values from a
blood sample measured on a YSI 2300 Stat Plus Analyzer.
• Number of Samples: 210
• Slope: 0.968
• Correlation Coefcient (r): 0.997
• y-intercept: 0.165 mmol/L
• Range tested: 0.5 - 12.4 mmol/L
These results indicate that Nova Lactate Plus Meter compared
well with the laboratory reference method.