56
Codes
Code Alerts Action
E-0 Software error detected Perform the test again. If you get the
same error, call Nova Technical Support.
E-1 Hardware error detected Perform the test again. If you get the
same error, call Nova Technical Support.
E-2 Meter temperature is outside Move the meter to an area
of the range for testing where the temperature is acceptable
(41°-113°F), allow meter to adjust to the
temperature. Repeat testing.
E-3 Defective or previously used Repeat test with a new
test strip is detected test strip.
E-4 A short sample (control Repeat test with a new test
solution or blood) is identied strip and adequate sample.
LO or If test results are outside of the Repeat test. If you get the
HI measuring range of the meter same result, consult your physician.
Battery level is low Replace the battery
Error Messages