Testing a Quality Control Solution
1. Grasp a test strip with logo side up and gold
side down. Then insert the gold end into the
meter. (See Meter gure.)
2. Identify the sample as a control; Use the left
or right arrow buttons to nd the de-
sired control level: QC1 or QC2 .
3. Shake the control solution well before each
use. Discard the rst drop.
4. Place a drop of control solution from the bottle
at the end of the test strip until the solution is
drawn into the well of the test strip.
5. A Lactate quality control test result is avail-
able on-screen in 13 seconds.