30
NOTE: Before running a blood or quality control test, check
that the expiration date of the test strips and control solution
has not elapsed. (See product insert sheets for details.)
The control solution test results should fall within the range
of results printed on the label of the control solution. You
should run a control solution test:
• Before using your Nova Lactate Plus Meter for the rst
time and at least once a week thereafter
• Each time you open a new box of Nova Lactate Test Strips
• If you leave the Nova Lactate Test Strip vial cap open
for an extended period of time
• If you drop your Nova Lactate Plus Meter
• If your results are higher or lower than expected
Testing: Blood and Quality Control Solutions