Precision
Precision of Nova Lactate Plus Meter was measured with both
venous blood samples and control solution in laboratory. The
results for the blood samples are shown below:
Summary of Lactate Plus Within-Run Precision - Blood Results
Lactate Blood Level
L1
(N=20)
L1
(N=20)
L1
(N=20)
L1
(N=20)
L1
(N=20)
Lot 1
Mean 1.6 6.5 10.8 18.1 22.1
CV% 3.6 1.9 1.6 2.7 3.2
SD (mmol/L) 0.06 0.12 0.17 0.49 0.70
Lot 2
Mean 1.2 6.1 10.3 16.9 20.7
CV% 4.9 2.4 1.6 3.9 4.9
SD (mmol/L) 0.06 0.15 0.17 0.66 1.01