Procedure 1
The Nova Lactate Plus Meter cleaning and disinfecting
procedure was validated a total of 10,950 times by Nova
Biomedical to simulate 3 year use-life of the Monitor. The
validation testing corresponded to cleaning and disinfecting
10 times a day for 3 years.
Acceptable Cleaning and Disinfecting Materials
Nova Biomedical recommends the use of Clorox Health-
care® Bleach Germicidal Wipes, EPA Registration #67619-
12, or any disinfectant product with EPA Registration
#67619-12 may be used.
Meter Cleaning and Disinfection
Clean and disinfect after each patient use by following this
protocol to help ensure effective cleaning and disinfec-
tion. Cleaning is not the same as disinfecting. Cleaning is