Interfering Substances
Creatinine Interferences:
The Nova StatSensor Xpress-i Creatinine Meter exhibits
no interference from the following substances up to the
following concentration levels:
Substances Conc
mg/dL(mmol/L)
Acetaminophen 10.0 (0.66)
Ascorbic Acid 3.5 (0.20)
Bilirubin 15.0 (0.26)
Cholesterol 1000.0 (25.8)
Creatine 4.0 (0.35)
Dopamine 10.0 (0.53)
Glucose 500.0 (27.8)
Hematocrit (RBC) 30%-60%
Heparin 120 units/dL
L-Dopa 300.0 (15.21)
Substances Conc
mg/dL(mmol/L)
D(+) Maltose Mono 100.0
Oxygen All Conc
Salicylate 30.0 (1.87)
Triglycerides 1000.0 (8.78)
Uric Acid 20.0 (11.4)