43
This section describes how to perform Linearity tests with the
StatSensor Xpress-i Creatinine Meter. There are 5 levels in the
StatSensor Creatinine Linearity kit.
Refer to the StatSensor Creatinine Linearity Kit
package insert sheet for complete instructions, indi-
cations, precautions, and limitations of the system.
1. Insert a test strip into the meter. All segments of screen will
display for 2 seconds. Then a ashing blood drop will display.
NOTE:Ifstripisre-
movedbeforethetest
startsorisnotused
forover2minutes,
thescreenwillgo
blank.
Testing a Linearity/Prociency Solution