53
Troubleshooting
Error Codes
There are 8 Error Codes to inform you of problems with the
meter. This section provides action procedures when these
Error Codes are displayed. The error code displays after the
test strip is inserted and the all segments screen displays for
2 seconds. If the beeper is enabled, there are also 3 quick
beeps. Then the Error Code is displayed on the screen.
E0 Software Error
A software error has been detected.
Action: Perform the test again. If you get
the same error again, remove and reseat
the battery. If the error continues, call Nova
Technical Support.