32
Testing: QC/Linearity/Prociency/Blood Samples
This section describes how to run QC (Quality Control),
Linearity Solution, Prociency Solution, and blood samples.
When to Perform a QC Test
The Nova StatSensor Xpress-i Creatinine Meter includes
several quality control mechanisms that detect errors due to
system failures and operator performance. External control
materials are available from Nova Biomedical for verifying the
integrity of the Nova StatSensor Xpress-i Creatinine Meter.
These StatSensor Creatinine Control Solutions consist of 3
levels of ready-to-use liquid controls. They are formulated at
clinically relevant levels. The controls can be used as part of a
laboratory quality control program. Run the controls according
to the procedure in Section 2.3 Quality Control Test.