User Manual TXLD0012
P/N T2-UM-E-ENG Rev 10 Novoxel.com Page | 41
After tip cleaning, clean the toothbrush using the medical grade Alcohol.
After cleaning the Tip, place the Handpiece in its cradle, without the DG. Soak a fiber-less wipe or
gauze in the medical grade Alcohol (70%) in order to clean the Distance Gauge thoroughly. Cleaning
should be performed using disposable gloves.
Return the DG to the Handpiece.
Heat cleaning of the selected Handpiece will now automatically commence. The process is 6:00
minutes long.
You may now select to perform cleaning of the second Handpiece as well. A countdown of 30 seconds
will be displayed, in which you may add the second Handpiece to the cleaning process by pressing on
its icon. If selected, you will then be requested to repeat the cleaning steps. The system will complete
the cleaning procedure by heating the Tip/s for the selected Handpiece/s.
If the second Handpiece was not selected, the heat cleaning process will continue to complete the
process.
The following screen will be displayed:
Tip Heat Cleaning is Performed; A 6:00 Minute Countdown is Displayed
After 6:00 minutes, the cleaning process has been concluded, the Operate screen will be displayed.
Returning to the Home screen will stop the cleaning process.