User Manual TXLD0012
P/N T2-UM-E-ENG Rev 10 Novoxel.com Page | 49
Criss-Cross Mode
Consider working in Criss-Cross mode in deeply wrinkled segments. In this technique, the pulses (single
pulse mode only) are performed in the same location but not in the same orientation, as depicted in the
following figure:
“Criss-Cross” Pulse Mode
In addition, do not overstretch the patient’s skin during treatment. In case of treatment of loose skin
areas, or areas which are not flat (such as back of hands), gently hold the skin straight in order to
flatten it. When working below the eyes (always maintain safe distance from the eyes), apply minimal
stretching by hand in order to avoid skin folding.
Position the Handpiece perpendicularly to the skin prior to activation, as demonstrated in the
following figure:
Criss-Cross may be performed only in Single pulse mode and by applying
relatively moderate parameters.
It is recommended that the treatment be performed with the patient lying down.
It is important to treat only flattened skin segments, in order to avoid skin
pinching, especially in thin skin patients.