User Manual TXLD0012
P/N T2-UM-E-ENG Rev 10 Novoxel.com Page | 37
Chapter 6: Caring For the Tips
Tip Cleaning
Prepare the Following:
A receptacle with Analytical Grade IPA (Isopropyl Alcohol; Propan-2-ol AR, 91% or higher).
A standard, clean, electrical rotating toothbrush (provided by Novoxel).
For the Distance Gauge cleaning only 70% Pharmacopoeia (medical) grade alcohol (70% Ethanol).
Cleaning should be performed using disposable gloves.
Press on the “Tip Cleaning” Button in the left upper corner of the Home Screen.
Tip Cleaning Screen
Temperature adjustment - The system will prepare the Tip for cleaning by either heating up to at least 80˚C
or cooling down below 200˚C. The following screen will be displayed:
Adjusting Tip temperature for Cleaning (Heating or Cooling, as Required)
All device components cleaning should be performed using disposable gloves.
Follow all safety instructions and storage for this material and its use, provided by its
supplier.