User Manual TXLD0012

Page | 36 Novoxel.com P/N T2-UM-E-ENG Rev 10
Displayed Screen Comments
This safety warning is displayed when the
Distance Gauge has been removed and the Tip is heated to high
temperatures. The notification will be eliminated only when the DG
is reconnected. In case the Distance Gauge has been removed, hold
the Handpiece carefully to avoid injury. Connect the DG to
proceed.
Do not remove the DG during an ongoing treatment session.
The Handpiece has expired. Connect a new Handpiece. In case the
Handpiece connection does not solve the problem, shutdown as
described in sub-section 4.5, wait for 15 seconds, and turn the
system back on.
This icon accompanies fatal error notifications which require the
operator to contact service:
E-01X
1
Handpiece Motion failure, please shut down, restart and try
again. If error persists call service.
E-02X
1
Handpiece heating failure, please shut down, restart and try
again. If error persists call service.
E-03X
1
Handpiece is heating up beyond the allowable range. Please
shut down, restart and try again. If error persists call service.
E-004 Handpiece not connected. Please disconnect Handpiece and
reconnect properly and then restart. If error persists, call service.

Table of Contents