Chapter 5: Troubleshooting
The following Table describes all notifications/error screens which may be displayed during the operation
of the Tixel2 device:
Item # Description
1.1
1.2
The Handpiece which was pre-selected in the Home screen, will be heated to working
temperature (400°C) and sterilized. When heated the icon will become orange and blink.
The selected Handpiece shall be indicated by the red dot in the following icon:
When Handpiece temperature have reached 385°C and sterilized, the icon will become
green, and blink until treatment parameters are selected (see indents 4, 5 & 6 further below
in this table).
After the treatment parameters have been selected, the Handpiece will become active; The
green icon will be displayed continuously and treatment can be performed.
Note: The other Handpiece will remain at hibernation temperature (200°C) indicated by
an orange icon.
Preparing the Second Handpiece for Operation
Press on the hibernated Handpiece icon. The second Handpiece will be heated to working
temperature and sterilized for 1:00 minute. During the heating process, the icon will
become orange and blink.
After the sterilization process has been completed, the second Handpiece will have a
blinking green icon but without the red dot.
Switching between Handpieces
Press on the icon of the second Handpiece. The Handpiece icon will have a red dot, as
demonstrated below:
The icon will blink until treatment parameters are selected (see indents 4, 5 & 6 further
below in this table). After the treatment parameters have been selected, the Handpiece will
become active; The green icon will be displayed continuously and treatment can be performed.
“X” stands for either 1) signifying the left Handpiece or 2) signifying the right Handpiece.