User Manual TXLD0012
Page | 62 Novoxel.com P/N T2-UM-E-ENG Rev 10
Check the integrity of the Handpiece’s and its connection to the console.
An audible “click” must be heard upon insertion of the connector to its inlet in the console.
In any case of malfunction or any abnormality, the device and/or Handpiece should be delivered to a
certified service provider for servicing.
Carefully remove the upper cover of the handpiece to make sure that the warranty seal below is not
broken.
When turning on the device listen to the sound of the blower. If there is an abnormal noise or no sound
at all, the system should not be operated. Deliver it to a certified service provider for servicing.
Verify that the white LED in the Handpiece is lit.
Handpiece maintenance after 250,000 pulses or 3 years operation.
Each Handpiece should be serviced by your local distributor after reaching 3 years of operation or
250,000 pulses.