14
Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
_ Drying
1
Remove moisture on the surface of the product using a dry
cloth, air gun, etc.
2
Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Lubrication (see “7 2 Lubrication”)
7 – 1 – 2 Automated Cleaning
Automated Cleaning with Cleaning Tools
Use the Cleaning Tools according to the following system figure.
*2: With O-ring (Order Code: D0312074080)
*3: With O-ring (Order Code: D0312150150)
Attachment
Cleaning Adaptor SM
Slim Attachment Hub
Cleaning Adaptor 1TSlim Tube
Cleaning Adaptor HM
*2
Attachment
Cleaning Adaptor WPD
*3
Attachment
Cleaning Drying
Cleaning Adaptor G Air Gun
Cleaning Adaptor Luer M Cleaning Tube 6.0/L50 Cleaning Tube 6.0/L650 Cleaning Adaptor Luer F
*1
Cleaning Holder
*1 Select the adaptor which matches the shape of the connection of the washer-disinfector.

Table of Contents

Other manuals for NSK Primado2

Questions and Answers: