English
After the tubes are correctly positioned, close the Pump Cover by
turning the Pump Cover Lever to the ‘CLOSE’ position. (180 degrees
counterclockwise) (Fig.6)
To remove, ‘OPEN’ the Pump Cover Lever then remove the tube.
<CLOSE Position>
Align
Fig.7
Tube Clamp
Bottle Cap
Irrigation
Tube Needle
Tube Cap
Fig.8 Fig.9 Fig.10
Fig.6
9
Mount the Coolant Solution Hanger Post onto the Holder on the Control
Unit. (Fig.7)
Close the Tube Clamp, between the Irrigation Tube Needle and the Irrigation Pump. (Fig.8)
Insert the Irrigation Tube Needle into the Bottle Cap. (Fig.9)
Open the Tube Cap to supply air into the bottle. (Fig.10)
Open the Tube Clamp.
1)
2)
3)
4)
4-7 Insertion of the Irrigation Tube
4-6 Mounting the Coolant Solution Hanger Post
CAUTION
CAUTION
Fig.5
• Make sure that the tube is securely
set on the rollers when closing
the Pump Cover. If the tube is not
correctly positioned on the Rollers
and the cover is closed, the tube
could be cut or sheared. (Fig.5)
• Do not mount the Irrigation Tube
except the person who specially
educated.
•
Do not mount the Irrigation Tube
during Power is ON.
• Do not operate the Irrigation Pump if the tube is bent or the Tube Clamp is in the closed position. This
could cause the tube to burst or slip out of the bottle.