4-8 Compatibility check of Internal Spray Nozzle/Drill
Internal Spray nozzles accompanied with this product; is not necessarily fitted into all the drills on the market.
Follow the instructions given below for confirmation prior to use.
Failure to do so or to fit the Internal Spray Nozzle into drills may cause a leakage of saline solution, which will result
in problems such as rust or sudden stop of equipment during use.
* To mount the Drill, refer to attached Handpiece Operation Manual.
Instructions:
Attach a bottle of saline solution to the Control Unit.
Connect the Internal Spray Nozzle into the tip of the irrigation tube.
Insert the Internal Spray Nozzle into the drill from the back. (Fig.11)
Purge at “Maximum” for 5 seconds.
1)
2)
3)
4)
Points to be checked:
4-9 Irrigation Nozzle Attachment
X-SG20L and SG20 has 3 irrigation methods available depending on tool and application; External, Internal or both,
for installation detail, refer to the attached handpiece Operation Manual.
4-10 Attaching the Tube Holder
Use the Motor Cord as a strain relief for the ‘Irrigation Tube’. It is
easier to insert Motor Cord first, then the Irrigation Tube. (Fig.12)
4-11 USB flash drive Socket (Surgic Pro
+
ONLY)
Open the Socket Cover on the Control Unit side.
Completely insert the USB flash drive. The USB mark will appear
on the LCD DISPLAY when USB flash drive is recognized by the
device. (Fig.13)
1)
2)
5. Setting
5-1 Clock and Date
Cleanliness of the saline solution coming out from the drill; if solution is colored there could be rust inside of the
drill. If so renew the drill.
Water Flow; if the flow is low and/or the flow from the drill is asymmetric, renew the drill.
No water leakage between Internal Spray Nozzle and drill before use, ensure no water is leaking from the entry
point of irrigation nozzle, a broken seal or no seal in the drill may be the cause. Replace the drill even if its new,
saline solution ingress into handpiece will cause malfunction.
By setting the date and time, data on treatment will be recorded by day. (See 6-5 USB Function to learn how to
store data on a USB flash drive.)
CAUTION
If malfunction such as a leakage of saline solution from the back of Contra Head is detected during
use, stop using and perform some troubleshooting.
CAUTION
Do not apply strong impact to USB flash drive when
it is connected. When the USB flash drive is not
connected, close the Insertion Opening Cover.