6-2 IMPLANT SYSTEM
6-3 Calibration Function
The rotation resistance of a handpiece depends on the handpiece model, condition, and internal wear of the
handpiece gears. This Control Unit incorporates an automatic function to recognize the level of the resistance of the
attached handpiece and motor.
6-3-2 Conducting Calibration
Fig.14
Fig.15
Surgic Pro Series can memorize 8 different IMPLANT SYSTEMs. Also, you can program 8 steps in one IMPLANT
SYSTEM number which you use. When you set the program, refer to values which Implant manufacture
recommends.
When programming each IMPLANT SYSTEM, the SYSTEM number may be changed by pressing the PROGRAM (+)
and (-) Key simultaneously.
Calibration Bur
Torque calibration
No-load calibration
torque display “L
Hold the motor in your hand and make sure that “L” is
displayed on the torque display. Push [CAL] key without
applying load. The calibration bur will begin to rotate.
Calibration is complete when there is a beep and “donE”
appears on the display.
Load calibration
torque display “H”
Make sure “H” is displayed on the torque display, plug the
calibration bur into the Calibration Load. Push [CAL] key
(The calibration is done while applying Calibration Load, so
securely grasp the handpiece. The calibration bur will begin to
rotate. Calibration is complete when there is a beep and “donE”
appears on the display).
1)
1-1)
1-2)
Attach Gear Ratio (Reduction Ratio) 20:1 handpiece to the motor.
Attach the Calibration Bur to the handpiece.
Press the [CAL] key for about 2 seconds. There will be a beep
and the display will switch to Calibration Mode and "CAL" is
displayed.
1)
2)
3)
6-3-1 Calibration preparation
NOTICE
NOTICE
• It is possible to choose programs that have been skipped using the Control Unit.
CAUTION
This equipment is optimized to obtain the highest calibration accuracy at a gear ratio of 20:1. For
calibration USE ONLY the 20:1 reduction handpiece listed on “12. Contra Angle Handpieces and
Accessories”.
• Calibration should be performed only on NSK handpiece.
Do not apply load during no-load calibration. If load is applied, FAILwill appear on the LCD, and
calibration will stop.