4) Release the Foot Control Pedal, and the Micromotor will automatically stop.
6-5 USB Function (Surgic Pro
+
ONLY)
The internal clock is used along with the different operations of this device (date used, starting time, length of time,
Gear Ratio,level of Coolant Solution Flow, Rotation Speed, FWD/REV, Torque, Program Number), making it possible
to store data in chronological order on USB flash drive (CSV files, BMP files).
Stored data can be viewed on a personal computer.
Fig.17
Fig.18
Torque
Time
Time
Date
Torque
Rotation Speed
Gear Ratio
Irrigation Volume
Program Number
*BMP File
Set torque index
NOTICE
• A BMP file can only be created only when Gear Ratio is 20:1 and the rotation speed is 50min
-1
or less.
Folders are created in
order by date
*Files and graphs saved
CSV file BMP file
15
CAUTION
• Using the Micromotor with light ‘ON’ for more than 3 minutes continuously, motor will become hot (The
maximum surface temperature might reach 45˚C by the situation used). In such cases, stop using the
Micromotor till the light part cools down, or use the Micromotor without turning on the light.
•
Activation of the torque limiter: During use, when the drilling load reaches the preset torque upper limit, the
integrated torque limiter automatically activates to prevent over torque. When the torque limiter activates,
the motor stops after beeping for 1 second. To reactivate the Micromotor, release the speed control pedal
and depress it again.
CSV file BMP file
NOTICE
• When the clock is not set, the time and date will not be recorded on data. It is recommended that the
clock be set before using the device.