*Autoclaving
Remove blood and debris from the handpiece and motor.
Clean inside the handpiece; by using the spray lubricant
Place those in autoclave pouch (not included in the package) and seal it.
Autoclavable up to max. 135˚C.
ex.) Autoclave for 20 min. at 121°C, or 3 min. at 134°C.
Keep the handpiece in the autoclave pouch to keep it clean until you use it.
* EN13060/EN ISO17665-1 recommends autoclaving for 3 minutes (minimum holding time) at 134°C or 15
minutes(minimum holding time) at 121°C.minutes(minimum holding time) at 121°C.
NSK recommends Class B sterilization.
1)
2)
3)
4)
5)
CAUTION
11. Sterilization
This handpiece and motor can be autoclavable up to Max.135°C.
• Autoclave sterilization is recommended.
• Autoclave sterilization is required for the first time use and after each patient as noted below.
CAUTION
Only Items Specified below can be autoclaved.
The following items can be autoclaved.
• Handpiece • Micromotor with Motor Cord • Handpiece Stand
• Internal Irrigation Nozzle • Tube Holder • Nozzle Holder
• Protection Plug • Calibration Bur
• Clean and lubricate the handpiece before autoclaving. Autoclaving a handpiece stained with blood or debris could
cause damage to the handpiece.
• Do not lubricate the micromotor.
• Do not disconnect the motor cord from the micromotor.
• The Irrigation tube is a single use and cannot be autoclaved.
• We do not recommend you to sterilize on plasma sterilization or EOG sterilization.